YOKOGAWA WT1600FC User Manual page 21

Impedance meter communication interface
Table of Contents

Advertisement

Explanation
IM 760151-17E
Carry out the following settings when using a controller to set information that can be
specified through key operation on the WT1600FC or when outputting setting
parameters or output waveform data to the controller.
Selecting the Baud Rate
Select the baud rate from the following.
1200, 2400, 4800, 9600, and 19200
Selecting the Data Format
Select the combination of data length, parity, and stop bit from the following.
8-NO-1, 7-EVEN-1, 7-ODD-1, and 7-NO-2
Selecting the Handshaking Method
Select the transmit data control and receive data control from the following.
NO-NO, XON-XON, XON-RTS, and CTS-RTS
Selecting the Terminator
Select the terminator from the following. The menu of the WT1600FC selects the
terminator that is used when transmitting data from the WT1600FC. Use "Lf" or
"Cr+Lf" for the terminator when receiving the data on the WT1600FC.
Cr, Lf, and Cr+Lf
2.6 Setting Serial Communications