YOKOGAWA WT1600FC User Manual page 89

Impedance meter communication interface
Table of Contents

Advertisement

[:INPut]:SCALing:STATe?
Function
Queries the scaling ON/OFF states of all
elements.
Syntax
[:INPut]:SCALing:STATe?
Example
:INPUT:SCALING:STATE? -> :INPUT:
SCALING:STATE:ELEMENT1 0;
ELEMENT2 0;ELEMENT3 0;ELEMENT4 0;
ELEMENT5 0
[:INPut]:SCALing[:STATe][:ALL]
Function
Collectively turns ON/OFF the scaling of all
elements.
Syntax
[:INPut]:SCALing[:STATe]
[:ALL] {<Boolean>}
Example
:INPUT:SCALING:STATE:ALL OFF
Description Scaling of all power measurement elements are
collectively set. To set the scaling on
impedance measurement elements, use the
"[:INPut]:SCALing[:STATe]:
ELEMent<x>" command. This command can
be used only on models that are equipped with
power measurement elements.
[:INPut]:SCALing[:STATe]:ELEMent<x>
Function
Turns ON/OFF the scaling of the element or
queries the current setting.
Syntax
[:INPut]:SCALing[:STATe]:
ELEMent<x> {<Boolean>}
[:INPut]:SCALing[:STATe]:
ELEMent<x>?
<x> = 1 to 5 (element)
Example
:INPUT:SCALING:STATE:ELEMENT1 OFF
:INPUT:SCALING:STATE:ELEMENT1? ->
:INPUT:SCALING:STATE:ELEMENT1 0
[:INPut]:SYNChronize?
Function
Queries the synchronization source of all power
measurement elements.
Syntax
[:INPut]:SYNChronize?
Example
INPUT:SYNCHRONIZE? -> :INPUT:
SYNCHRONIZE:ELEMENT1 I1;
ELEMENT2 I2;ELEMENT3 I3;
ELEMENT4 I4
Description This command is invalid on models that are only
equipped with impedance measurement
elements, since the mode is fixed to impedance
measurement.
IM 760151-17E
[:INPut]:SYNChronize[:ALL]
Function
Collectively sets the synchronization source of
all power measurement elements.
Syntax
[:INPut]:SYNChronize[:ALL] {U<x>|
I<x>|NONE}
<x> = 1 to 4 (power measurement element)
NONE = No synchronization source
Example
:INPUT:SYNCHRONIZE:ALL I1
Description This command is invalid on models that are only
equipped with impedance measurement
elements, since the mode is fixed to impedance
measurement.
[:INPut]:SYNChronize:ELEMent<x>
Function
Sets the synchronization source of the power
measurement element or queries the current
setting.
Syntax
[:INPut]:SYNChronize:
ELEMent<x> {U<x>|I<x>|NONE}
[:INPut]:SYNChronize:ELEMent<x>?
<x> = 1 to 4 (power measurement element)
NONE = No synchronization source
Example
:INPUT:SYNCHRONIZE:ELEMENT1 I1
:INPUT:SYNCHRONIZE:ELEMENT1? ->
:INPUT:SYNCHRONIZE:ELEMENT1 I1
Description This command is invalid on models that are only
equipped with impedance measurement
elements, since the mode is fixed to impedance
measurement.
[:INPut]:VOLTage?
Function
Queries all settings related to the voltage
measurement.
Syntax
[:INPut]:VOLTage?
Example
:INPUT:VOLTAGE? -> :INPUT:VOLTAGE:
RANGE:ELEMENT1 1.0000E+03;
ELEMENT2 1.0000E+03;
ELEMENT3 1.0000E+03;
ELEMENT4 1.0000E+03;
Description This command is invalid on models that are only
equipped with impedance measurement
elements, since the mode is fixed to impedance
measurement.
[:INPut]:VOLTage:AUTO[:ALL]
Function
Collectively turns ON/OFF the voltage auto
range of all power measurement elements.
Syntax
[:INPut]:VOLTage:AUTO
[:ALL] {<Boolean>}
Example
:INPUT:VOLTAGE:AUTO:ALL ON
Description This command is invalid on models that are only
equipped with impedance measurement
elements, since the mode is fixed to impedance
measurement.
5.10 INPut Group
5
5-53

Advertisement

Table of Contents
loading

Table of Contents