:STATus:ERRor?
Function
Queries the error code and message
information (top of the error queue).
Syntax
:STATus:ERRor?
Example
:STATUS:ERROR? ->
113,"Underfined Header"
Description • When there is no error, "0, "No error"" is
returned.
• The message cannot be returned in
Japanese.
• You can specify whether or not to add the
message using the "STATus:QMESsage"
command.
:STATus:FILTer<x>
Function
Sets the transition filter or queries the current
setting.
Syntax
:STATus:FILTer<x> {RISE|FALL|BOTH|
NEVer}
:STATus:FILTer<x>?
<x> = 1 to 16
Example
:STATUS:FILTER2 RISE
:STATUS:FILTER2? -> :STATUS:
FILTER2 RISE
Description Specify how each bit of the condition register is
to change to set the event. If "RISE" is specified,
the event is set when the bit changes from "0" to
"1."
:STATus:QENable
Function
Sets whether or not to store messages other
than errors to the error queue (ON/OFF) or
queries the current setting.
Syntax
:STATus:QENable {<Boolean>}
:STATus:QENable?
Example
:STATUS:QENABLE ON
:STATUS:QENABLE? -> :STATUS:
QENABLE 1
:STATus:QMESsage
Function
Sets whether or not to attach message
information to the response to the
"STATus:ERRor?" query (ON/OFF) or queries
the current setting.
Syntax
:STATus:QMESsage {<Boolean>}
:STATus:QMESsage?
Example
:STATUS:QMESSAGE ON
:STATUS:QMESSAGE? -> :STATUS:
QMESSAGE 1
IM 760151-17E
:STATus:SPOLI? (Serial Poll)
Function
Executes serial polling.
Syntax
:STATus:SPOLl?
Example
:STATUS:SPOLL? -> :STATUS:SPOLL 0
Description This is a command specific to the serial (RS-
232) interface. An interface message is
available for the GP-IB interface.
5.15 STATus Group