5.5.2

Flow Connections Overview

UHPLC System
Operating Manual
The picture illustrates the liquid flow path through the system:
Figure 22: Flow connections in the Vanquish system (example)
The Vanquish system has been designed for optimized and simple leak
liquid and waste handling:
Leak liquid from the solvent rack, detector, autosampler, and pump
flows through pipes from the modules to the system base drain port.
Seal wash and needle wash liquids from the pump and autosampler,
as well as condensate from the autosampler, are also guided to the
system base drain port.
Leak liquid from the column compartment flows to the column
compartment drain port.