6.2

Safety Guidelines for Operation

UHPLC System
Operating Manual
When operating the system, pay attention to the following safety
guidelines:
Observe all warning messages and precautionary statements presented
in
Safety Precautions
(} page 19).
Observe all warning messages and precautionary statements presented
in the Safety Guidelines for Operation section in the Operating Manual
for the individual modules of the Vanquish system.
NOTICE
Pay attention also to the following guidelines:
When operating the chromatography system, always set the lower
pressure limit for the pump. This prevents damage resulting from
leakage or from running the pump dry.
If there is evidence of leakage in the device, turn off the pump flow
and remedy the situation immediately.
If the pump flow is interrupted, act appropriately to protect the
components in the detector. For details, refer to the Operating
Manual for the detector.
Always verify that the autosampler is turned on before the pump
flow is on and pressure builds up. If the autosampler is turned off,
for example, after a power failure, stop the pump flow and wait until
the pressure is down to zero before turning on the autosampler or
other modules again.