Maintenance and Service
Removing the System
Modules
Page 110
1. Slide the carrying handles that were installed when the system was
shipped on the rails at the right and left of the module from the rear
as shown in the figure below.
The final position of the handles must match the foam spacers used
for transport (about 15 cm from the rear of the module to the
center of the handle).
Figure 35: Sliding the carrying handle on the left rail
2. Fix the attachment screws on the carrying handles as shown below.
Figure 36: Carrying handles attached
No.
Description
1
Carrying handles
2
Attachment screw (one on each carrying handle)
CAUTION—Risk of Injury and Device Damage
Handles can slide off the device if the attachment screws are not fixed
properly. Also, the slide-in module can fall out if not fixed properly. This
could cause personal injury and damage to the device. Verify the
following before lifting a device:
The handles are fixed properly and cannot slide off the device.
The four screws on the slide-in module are tightened and the slide-in
module cannot fall out (see Operating Manual of the detector, for
example).
3. Lift the front part of the module by using the carrying handles.
Remove the module from the stack towards the front and carefully
place it on a stable surface.
UHPLC System
Operating Manual