5.4.5

System Base Lock

5.4.5.1
Unlocking the System Base
UHPLC System
Operating Manual
A lock is integrated into the system base. Unlock the system base if you
want to move the system base and lock it if you want to prevent it from
being moved on the workbench. If your system configuration includes
one column compartment, you can switch between the moveable and
stationary mode using the system base locking tools shipped with the
system base.
Figure 17: System base locking tool
If your system includes two or three column compartments or a Charger,
a Thermo Fisher Scientific service engineer must dismount the
mentioned modules before you can (un)lock the system base with the
locking tools.
1. Have all but one column compartment and the Charger dismounted
by a Thermo Fisher Scientific service engineer.
2. Insert one locking tool into another.
Figure 18: Extended locking tool
3. Insert the combined locking tools into the lock port on the side of
the system base where the column compartment is installed.
If the column compartment is installed on the right side, insert the
combined locking tools into the right lock port as shown in the figure
below.