Decontamination
Parts required
UHPLC System
Operating Manual
NOTICE
Wipe up all liquids spilled onto the system immediately. If surfaces are
exposed for longer periods, these liquids can cause damage.
Decontamination is required, for example, when leakage or spillage has
occurred, or before service or transport of the system. Use a suitable
cleaning detergent or disinfectant to ensure that the treatment renders
the system safe to handle.
Suitable cleaning detergent (or disinfectant)
Purified water
Lint-free cloths or wipes
CAUTION—Explosive Gas Mixtures from Alcoholic Cleaning Detergents
Alcohol-containing cleaning detergents may form flammable and
explosive gas mixtures when exposed to air.
Use such cleaning detergents only when required and only in
adequately ventilated rooms.
Avoid open flames or exposure to excessive heat during the cleaning
process.
Wipe the cleaned components thoroughly dry after cleaning. Do not
operate the device before it is completely dry.
NOTICE
Observe the following:
Only use cleaning detergents that will not damage the surfaces of
the system.
Never use sharp tools or brushes for cleaning any surfaces.
Do not use sprays for cleaning.
Prevent cleaning detergent from entering the flow path.
Do not use excessively wetted cloth or wipes for cleaning. Prevent
any liquids from entering the functional components of the device.
Liquids can cause a short circuit when getting in contact with the
electronic components.
7
Maintenance and Service
Page 107