9
Specifications
9.1

System Specifications

Type
Materials in the flow path
Biocompatibility
Control
Good Laboratory Practice (GLP)
features
Safety features
Range of use
Ambient operating temperature
Ambient storage temperature
Ambient operating humidity
Ambient storage humidity
Operating altitude
Pollution degree
Emission sound pressure level
System dimensions
(height x width x depth)
System weight
With one column compartment,
without Charger
Page 130
The physical conditions of a standard system are specified as follows:
Specification
For materials in the flow path, refer to the Operating Manuals for the
modules in your system.
For solvent and additive information, see
Information
Core system in the standard
configuration: No
System base power button
Chromeleon 7
The system can be operated also with other data systems. For details,
contact the Thermo Fisher Scientific sales organization.
Predictive Performance functions for scheduling maintenance
procedures based on the actual operating and usage conditions of the
system.
All system parameters logged in the Chromeleon Audit Trail.
Leak detection and safe leak handling
Indoor use only
5 °C - 35 °C
-20 °C - 45 °C
20% - 80% relative humidity, non-condensing
Maximum 60% relative humidity, non-condensing
Maximum 2000 m above sea level
2
System with one column compartment, without Charger:
Typically approx. 55 dB(A), max. approx. 75 dB(A)
System with one column compartment, without Charger:
82 cm x 55 cm x 62 cm
Vanquish Core
Approx. 80 kg
(} page 25).
Horizon/Flex system in the
standard configuration: Yes
Vanquish Flex:
Approx. 90 kg
Solvent and Additive
Vanquish Horizon:
Approx. 110 kg
UHPLC System