Monitoring The Baseline - Thermo Scientific thermoscientific Vanquish UHPLC System Operating Manual

Table of Contents

Advertisement

6
Operation
6.6.3

Monitoring the Baseline

Page 94
3. On the ePanel Set, click the tab for a module.
4. Use the controls (for example, buttons, sliders) to perform
commands.
5. If a function that you want to perform is not available on a panel,
press the F8 key to open the Command window. From this window,
you can access all commands that are available for the system.
TIP
The commands and parameters that are available in the Command
window may vary, depending on the user level in the Command window
(Normal, Advanced, or Expert). To change the user level, right-click in
the commands list and select a different user level.
When an instrument is idle, you can monitor the baseline signals
without starting a sequence. During baseline monitoring, the monitored
signals are displayed on the real-time signal plot on the ePanel for the
module. You may have to add the signal to the plot in the Plot Properties
(for details, refer to the Help).
To monitor the baseline signals:
1. Open the ePanel Set.
2. On the toolbar above the ePanel Set, click Monitor Baseline.
3. In the Select Channels to Monitor dialog box, select the signals that
you want to monitor.
4. To stop monitoring the baseline, click Stop on the toolbar.
TIP
Monitor baseline data is overwritten each time baseline monitoring is
started. If you want to save the data permanently, define the location in
the Monitor Baseline Save Preferences dialog box. For details, refer to
the Help.