Low Level Alarms (2.8.2.) - Siemens SITRANS LUT400 Operating Instructions Manual

Ultrasonic controllers
Hide thumbs Also See for SITRANS LUT400:
Table of Contents

Advertisement

High Level Value OFF (2.8.1.3)
Sets the material level (defined in Units (2.1.1.) (Page 168)) at which the High Level Alarm
will de-activate.
Values
Assigned Relay (2.8.1.4)
Determines which relay (if any) will be activated when the High Level Alarm activates.
Options
Alarm State (2.8.1.5)
Read only. Used to view the current state of the High Level Alarm.
Options
9.2.8.2

Low Level Alarms (2.8.2.)

Reported when material level is within a user-defined range (see Low Level Value ON
(2.8.2.2.) (Page 197) and Low Level Value OFF (2.8.2.3.) (Page 198)).
Enable (2.8.2.1.)
Enables/disables Low Level Alarm.
Options
Low Level Value ON (2.8.2.2.)
Sets the material level (defined in Units (2.1.1.) (Page 168)) at which the Low Level Alarm
will activate.
Values
SITRANS LUT400
Operating Instructions, 07/2019, A5E33329501-AC
Range: 0.000 to 99999.000
Default: 0.000
*
ACTIVE
INACTIVE
*
Range: 0.000 to 99999.000
Default: 0.000
NO RELAY
RELAY 1
RELAY 2
RELAY 3
ENABLED
DISABLED
197