Confidence (3.2.9.2.)
Indicates echo reliability: higher values represent better echo quality. The display shows the
echo confidence of the last measurement. Echo Threshold (2.12.2.2.) (Page 220) defines the
minimum criterion for echo confidence.
Values (view only)
In SIMATIC PDM, open the menu Device – Echo Profile Utilities and click on the tab Echo
Profile.
Echo Strength (3.2.9.3.)
Displays the absolute strength (in dB above 1 μV rms) of the echo selected as the
measurement echo.
Values (view only)
In SIMATIC PDM, open the menu Device – Echo Profile Utilities and click on the tab Echo
Profile.
Noise Average 3.2.9.4.)
Displays the average ambient noise (in dB above 1 μV rms) of a noise profile after each
measurement.
The noise level is a combination of transient acoustic noise and electrical noise (induced into
the transducer cable or receiving circuitry). See Noise Problems (Page 276).
Noise Peak (3.2.9.5.)
Displays the peak ambient noise (in dB above 1 μV rms) of a noise profile after each
measurement.
9.3.3

Maintenance (3.3.)

Note
Default settings in the parameter tables are indicated with an asterisk (*) unless explicitly
stated.
9.3.3.1

Transducer Enable (3.3.1.)

Enables/disables transducer from taking measurements.
Options
SITRANS LUT400
Operating Instructions, 07/2019, A5E33329501-AC
Range: -20 to 128
Range: -20 to 128 dB
*
ENABLED
DISABLED