Siemens SITRANS LUT400 Operating Instructions Manual page 66

Ultrasonic controllers
Hide thumbs Also See for SITRANS LUT400:
Table of Contents

Advertisement

Commissioning
6.3 Quick Start Wizards
Start of QS Volume Wizard
Note
The introduction screen is displayed only on the device when using the local push buttons.
This screen is not part of the Quick Start when using SIMATIC PDM.
Shows the type of Wizard to be executed.
Options
Transducer
Specifies the Siemens transducer connected to the device.
Options
Temperature Source
Source of the temperature reading used to adjust the speed of sound.
Options
See Temperature Source (2.12.1.3.) (Page 218) Temperature Source for more details.
Fixed Temperature
Use this feature if a temperature sensing device is not used.
Value
This parameter is only displayed if FIXED TEMPERATURE selected for Temperature
Source.
64
CANCEL, START
NO TRANSDUCER, XRS-5, XPS-10, XPS-15, XCT-8, XCT-12, XPS-30, XPS-40,
XLT-30, XLT-60, STH
Default: NO TRANSDUCER
TRANSDUCER, FIXED TEMPERATURE, EXTERNAL TS-3, AVERAGE OF SENSORS
Default: TRANSDUCER
Range: -100.0 to +150.0 °C
Default: +20.0 °C
Operating Instructions, 07/2019, A5E33329501-AC
SITRANS LUT400

Advertisement

Table of Contents
loading

Table of Contents