Read this section of the manual before beginning installation of the 5000 Analyzer System. Fail-
ure to do so, and /or use of the Analyzer in a manner not specified in this manual, may impair the
!
protection against fire, electrical shock, and personal injury originally provided by this equipment.
WARNING
In addition, failure to follow the installation and start-up instructions may void the instruments
warranty.

Prior to Installation

The 5000 Field Unit is approved for the following hazardous area classifications: Class I; Group B,
C, D; Division 1 as defined by the U.S. National Electrical Code (NEC) and CENELEC EEx d IIC T6.
The 5000 Analyzer System is typically installed with the field unit near the sample tap (normally
in a hazardous area) and the 5000 controller located in a control room. The controller is suitable for
installation in NEC Class 1 ; Groups A, B, C & D; Division 2 areas. The field unit and controller may
be separated by up to 600 meters (2000 feet). Prior to installation a site plan should be prepared de-
fining the mounting locations of the field unit, controller, and sample system components. The plan
should also detail the AC input(s), interconnect wiring, and sample line routing.

Unpacking and Inspection

Remove components from the packing case(s) carefully; check contents against packing list. Inspect
all components for obvious damage and broken/ loose parts or fittings. Notify the carrier and AM-
ETEK Service (1-800-537-6044) immediately if parts are missing or damage is found.

Field Unit Installation

Figure 4 shows installation dimensions for standard (Division 1 and 2) and CENELEC versions of
the field unit. Locate the field unit as close as possible to the sample tap and bolt into place. The unit
should be protected from direct exposure to weather and located so that the ambient temperature
specifications will not be exceeded.

Controller Installation

Mount the controller indoors in desired location and bolt in place. Figures 5a through 5d show in-
stallation dimensions for all controller mounting configurations.

Electrical Connections

Refer to the Customer Drawing (CD) package accompanying this manual for 5000 System Inter-
!
connect Wiring Configuration and Figure 7 for 5000 Controller back plane terminal locations.
WARNING

INSTALLATION

Installation 2-1