Passwords

Highlight the password to be viewed/altered; press
when the "New Value?" prompt is displayed. Press
words Menu; press
CANCEL
tion on the various security levels.
Maint_Pswrd
Confg_Pswrd
Setting both the Level 1 and Level 2 passwords to "0000" disables low level security. Associated
menus and system values may be accessed with no password required; press
NOTE
prompt is displayed. The Level 3 password cannot be altered.

Timing

Scroll through the following list of menu choices until desired item is highlighted; press
select. Enter the new timer value when the "New Value?" prompt is displayed. Press
change and return to the Configure Menu; press
AutoCalCycle
If cal cycle is to be run less frequently than once a day, enter the number of
days to elapse until the next cal cycle. This timer is used in conjunction with
Cal_Time. Cal_Timer should be set to 000000.
Cal_Timer
Countdown time in seconds between cal cycles. When this timer reaches zero
(times out), a verification cycle will be initiated. Enter the new timer value in
seconds when the "New Value?" prompt is displayed. Note that value entered
must be in six digit form, using lead zeros if necessary. Cal_Time should be
set to 00:00.
Sample_Cycles Number of complete cycles analyzer will remain in sample mode for each
sample point. Press cancel to roll through the variables SamplCycl_X where
X = 1
4. Sample cycle 5 corresponds to the complete verification cycle.
Equil_Cycles
Number of complete cycles to allow for system equilibration before each sample
point measurement; and before and after verification. Press cancel to roll
through the variables EquilCycl_X where X = 1
4 correspond to the equilibration period before sample points 1 through 4 are
analyzed. EquilCycle 5 corresponds to the equilibration period before and
after verification.
Avg_Cycles
Number of complete cycles used to average data when analyzer is configured
to use enhanced noise smoothing. Refer to Flags2 in this section for more
information on configuring the analyzer to use enhanced noise smoothing.
Enhanced noise smoothing will affect analyzer response time.
NOTE
to quit. Refer to Page 3-1: System Security and Passwords for informa-
View/alter the Maintenance Level (Level 1) password.
View/alter the Configure Level (Level 2) password.
5. SamplCycl_1 through 4 correspond to sample points 1 through
to select. Enter new four-digit password
ENTER
to save change and return to the Pass-
ENTER
to quit.
CANCEL
when password
ENTER
to
ENTER
to save
ENTER
5. EquilCycl_1 through
Software Configuration 4-3