Display (Display Flags); Analog (Analog Flags) - Ametek 5000 Instrument Manual

Moisture analyzer with multi-point analysis
Hide thumbs Also See for 5000:
Table of Contents

Advertisement

Constants
Dry_CF
This parameter, Dry Cell Frequency, is the nominal cell frequency determined during
factory calibration of the instrument and is given for informational purposes only.
Cell_SN
Number assigned to a particular cell calibration. May be updated in configuration
data file upon replacement.
MoistGen_SN
A Serial number assigned to a calibrated Moisture Generator. May be updated in
configuration data file upon replacement.

Display (Display Flags)

This menu item is used to select the system values to be continuously displayed on the first three
lines of the display screen. Select which line is to be altered by highlighting Line_X_Flag where
X = 1
3. Enter the flag value from the following list when the "New Value?" prompt is displayed.
Press
to save change and return to the Configure Menu; press
ENTER
noted, the same flags are used for the analog output and alarm flags (below).
Display Flag Values
0 None
1 Point 1 - Held Moisture During Equilibration, Cal, and Zero
2 Point 2 - Held Moisture During Equilibration, Cal, and Zero
3 Point 3 - Held Moisture During Equilibration, Cal, and Zero
4 Point 4 - Held Moisture During Equilibration, Cal, and Zero
5 Roll Through Points (Display Last Measurement)
6 Current Moisture (Display Current Point and Moisture)
7 Cell Frequency (Cell_Freq) in Hz
8 Delta Frequency (Delta_Freq) in Hz
12 Selected Streams
14 Status of Valve and Alarm Relays
15 Period Timers & Cal_Time OR Cal_Timer
(See
in Page 4-3: Timing)
NOTE
16 Output Range (Channel 1)
* These displays can be altered by the user. Refer to Page 4-2: Labels for more information on
changing display formats.
** These flags are for display lines only, they do not apply to alarm or analog output flags
Analog (Analog Flag)
This menu item is used to select the system values to be linked to the four 4 to 20 mA analog out-
puts. Select the output flag to alter by highlighting OutputX_Flag where X = 1 - 4. Enter the new
flag value when the "New Value?" prompt is displayed. The values to be entered are identical to Dis-
play Flags 1 through 11. Press
to quit.
4-2 5000 Multi Point Moisture Analyzer
to save change and return to the Configure Menu; press
ENTER
to quit except where
CANCEL
Example of Display
Blank
* 1.000 ppm Vol H2O Pt1
* 1.000 ppm Vol H2O Pt2
* 1.000 ppm Vol H2O Pt3
* 1.000 ppm Vol H2O Pt4
1.000 ppm Vol H2O PtX (X=1
1.000 ppm Vol H2O PtX (X=1
1200 hz Cell Freq
0.50 hz Delta Freq
** Streams Slctd 1234
** 010-Alr 00001010 Vlv
** 27-Ref 08:00-Cal
or 27-Ref 003599-Cal
** 100.000 Rcdr Range
4)
4)
.
CANCEL