The following labels may be altered:
Alter Moist1 Lbl (Line_X_Flag = 1: Point 1 Held Moisture During Equilibration, and Cal)
Default: ~~~ ppm Vol H2O Pt1
Alter Moist2 Lbl (Line_X_Flag = 2: Point 2 Held Moisture During Equilibration, and Cal)
Default: ~~~ ppm Vol H2O Pt2
Alter Moist3 Lbl (Line_X_Flag = 3: Point 3 Held Moisture During Equilibration, and Cal)
Default: ~~~ ppm Vol H2O Pt3
Alter Moist4 Lbl (Line_X_Flag = 4: Point 4 Held Moisture During Equilibration, and Cal)
Default: ~~~ ppm Vol H2O Pt4

Constants

Use the arrow keys to scroll through the following list of system constants until item to be altered is
highlighted; press
ENTER
displayed. Press
ENTER
Convr_Factor Constant used to calculate moisture concentration. This value is factory set
and should not be changed unless factory service is performed.
Moist_Span

Flags2 (Configuration Flags)

The following list of software configuration features can be selected by altering the value of the vari-
able Flags2. The Flags2 value is essentially a set of positions for eight bits, or one byte, of computer
logic. Each bit position has a specific binary value when turned ON; all bits have a value of zero
when turned OFF. Similar to Flags1, the value entered for the Flags2 is the sum of binary values for
all of the features to be selected. Enter new flag value when the "New Value?" prompt is displayed.
Press
to save change and return to the Configure Menu; press
ENTER
Bit Value
1
Used for special applications only
2
Used for special applications only
4
Use enhanced noise smoothing (Boxcar Averaging). Note that a value
must also be assigned to the timer Avg_Cycles to use this feature
8
Used for special applications only
to select. Enter the new constant value when the "New Value?" prompt is
to save change and return to the Constants Menu; press
The constant cell verification factor; refer to Chapter 3: Verification.
Function
to quit.
CANCEL
to quit.
CANCEL
Software Configuration 4-7