Ametek 5000 Instrument Manual page 17

Moisture analyzer with multi-point analysis
Hide thumbs Also See for 5000:
Table of Contents

Advertisement

SPECIFICATIONS
Ranges
• Calibrated 0 to 1000 parts per million by volume (ppmv)
• Display continues to provide trend information above 1000 ppmv.
• Output capability in lbs/MMscf and dewpoint.
Outputs
• 80 character vacuum fluorescent display
• Two 4 to 20 mA dc analog signals with completely user settable
span; one analog output is auto-ranging
• RS-485 bi-directional serial port
Alarm Contacts
System Alarm
Computer Operating Properly (COP) alarm
Four Dedicated Concentration Alarm Contacts
Range ID (Channel 1 only)/Data Valid Contacts
(Contacts rated 30Vac/60Vdc, 50 VA maximum, 1A maximum)
Concentration Alarm Relays
Mechanical, AC/DC
Voltage 240 maximum
Power 50 VA maximum
(ex: 2A at 240 Vac)
Current Fused at 1A
Sample Switching Relays
Solid-State, AC only:
Voltage 250 maximum
Power 50 VA maximum
Current Fused at 1A
Sensitivity
0.02 ppmv (9.5 x 10
lb/MMscf)
–2
Repeatability
0.02 ppmv ( 9.5 x 10
lb/MMscf) or 1% of reading,
–2
whichever is greater.
(0.02 ppmv over 0 to 5 ppmv with optional super-activated dryer on
certain applications).
Accuracy
1.0 ppmv (0.05 lb/MMscf) or 5% of reading, whichever is greater
(0.1 ppmv over 0 to 5 ppmv with optional super-activated dryer on
certain applications).
Moisture Generator Value
Standard:
20 ppmv typical ± 10%
Low Option: 3 ppmv typical ± 10%
Operating Pressure
103 kPa gauge (15.0 psig)
Allowable Inlet Pressure
207 to 690 kPa gauge (30 to 100 psig)
Gas Flow Requirements
250 mL/min each at 103 kPa gauge (15 psig) for sample, reference,
and moisture generator
*Unless ordered as an extended range configuration.
Overview 1-7