have not been changed. Note that it is normal to see a slight change in values for
PolyCoeff_D and Moist_Span as these may be reset during zero and calibration
cycles.
TEST
CONFIG
To view configuration values, press
CANCEL
data formats will be displayed in the order shown.
• Select Streams
Used to select streams to be analyzed. If all four streams are currently selected,
the message "Streams Slctd 1 2 3 4" will be on line 2 of the display. In addition,
the stream currently being analyzed will be blinking. Enter the point number (1
Õ 4) to toggle. For example, if you want to de-select sample point 3 from normal
sample switching, press 3. Line 2 will read "Streams Slctd 1 2 4". Pressing 3 again
will re-select it. Press
• Set Time of Day
CANCEL
ENTER
At "Enter New Time" prompt, enter four digit time of day using the 24 hour clock
convention. Note that it is not necessary to enter punctuation between hours and
minutes.
Moist_Span
Moist_Gen
Convr_Factor
Flags_1
Flags_2
• Set Date
At "Enter New Date" prompt, enter six digit date in the form MM/DD/YY. Note that it is not nec-
essary to enter punctuation between month, day, and year.
• Set Auto Cal Time
At "Enter New Time" prompt, enter the time of day for the auto-calibration cycle to be run (this
becomes the value of the variable Cal_Time). Enter four digit time using the 24 hour clock conven-
tion. Note that it is not necessary to enter punctuation between hours and minutes. Cal_Time is
used in conjunction with Auto_Cal_Cycle to set the interval in days between calibration cycles.
Refer to Page 4-3: Timers for more information on setting cycle timers.
• System Tests Menu
The 5000 Controller has numerous self-test routines which can be run periodically as part of regu-
lar maintenance or used to assist in troubleshooting the analyzer. The system tests are discussed
in detail in Chapter 5.
• Configure Menu
Both Level 2 and Level 3 Configure Menus are accessible through this menu item. Refer to Chapter
5: Software Configuration for more information on the Configure Menus and associated func-
tions.
3-14 5000 Multi Point Moisture Analyzer
to return to the
/
TEST
CONFIG
to save changes and quit.
CANCEL
Line_1_Flag
Line_2_Flag
Line_3_Flag
Output1_Flag
Output2_Flag
Output3_Flag
to display the next parameter; press
ENTER
Main Menu. The following parameters and
Output4_Flag
PolyCoeff_5
PolyCoeff_0
PolyCoeff_A
PolyCoeff_1
PolyCoeff_B
PolyCoeff_2
PolyCoeff_C
PolyCoeff_3
PolyCoeff_4