Ametek 5000 Instrument Manual page 58

Moisture analyzer with multi-point analysis
Hide thumbs Also See for 5000:
Table of Contents

Advertisement

5) Set timers to achieve sample switching frequency of ≈ 10 minutes including equilibration
periods between sample analysis and to set duration of calibration cycle to 10 minutes. Press
the test/config key; select Configure Menu-timing-Sample_Cycles; set the following timers to
8 cycles.
Return to Timing menu; select Equil_Cycles; set the following timers to 2 cycles.
4-10 5000 Multi Point Moisture Analyzer
Alarm_1_Hi
60
Alarm_1_Low
0
Alarm_2_Hi
15
Alarm_2_Low
0
SamplCycl_1
SamplCycl_2
SamplCycl_5
EquilCycl_1
EquilCycl_2
EquilCycl_5
AMETEK 5000 Controller
with Multi-Point
Default System Security Passwords
Level 1:
Level 2:
Level 3:
1 1 1 1 [ENTER]
2 2 2 2 [ENTER]
.3 - 18 [ENTER]