4. If necessary, adjust R26 one turn for each 1000-ohm (≈1°C) difference between the specified and
actual resistances. (Turn clockwise to decrease resistance.)
5. Remove the DVM and connect the moisture generator plug to J2.
6. Re-install the lid.
7. Apply power to the system and allow it to warm up.
8. If necessary, repeat the procedure until no further adjustment is required to maintain 60°C.
6-18
5000 Multi Point Moisture Analyzer