Troubleshooting and Maintenance
!
WARNING

System Alarms

The system alarm contacts are activated when a condition exists which makes the analyzer output
questionable. The following error messages will be displayed on the bottom line of the analyzer display
when a system alarm condition exists:
CELL TEMP ALERT
LOW CELL FREQUENCY
HIGH CELL FREQUENCY
NEGATIVE DELTA FREQUENCY
GEN TEMP ALERT
DISPLAY OVER-RANGED
MEMORY IS CORRUPTED
It is common for some of these system alarms to activate momentarily during normal operation, espe-
cially during equilibration periods between measurement cycles. If system alarm persists, there may
be a problem with the measuring cell, the moisture generator, or the thermal controller.
Computer Operating Properly Alarm
This alarm, also known as the COP or watchdog alarm, activates when normal CPU logic flow is dis-
rupted. Possible triggers of this alarm are power failure, internal power failure, software problems, or
extreme electronic interference. When activated, the COP alarm generates a system reset but software
configuration parameters are retained. The system does not have to be reconfigured.
Any troubleshooting or maintenance to this Analyzer that requires access to the
interior of the various components must be performed by qualified service person-
nel only. There are no operator level serviceable parts, expendables, or maintenance
needs inside this equipment.
Moisture Cell temperature control is suspect (signal sent from
field unit).
Activates when measured cell frequency is < 500 Hz.
Activates when measured cell frequency is > 3500 Hz.
Activates when delta frequency is < –1 Hz.
Moisture Generator temperature control is suspect (signal sent
from field unit).
CAL ERROR
Activates when the new calculated span value (Moist_Span)
exceeds the limits Max_Cal_Err or Max_Span Refer to Part III:
Alarm Button for more information on these limits. Deactivates
after successful calibration.
A displayed value exceeds the allotted display area.
The RAM or RAM battery has failed causing factory default
parameters to be loaded into memory.
Troubleshooting and Maintenance 6-1