Kongsberg Simrad EK80 Installation Manual page 97

Hide thumbs Also See for Simrad EK80:
Table of Contents

Advertisement

Cable
Type
C16*
DC power cable
The DC power cable from the WBT power supply to the transceiver is provided with the
power supply. The Wide Band Transceiver (WBT) can be powered by a large capacity
battery. The battery must be placed close to the Wide Band Transceiver (WBT) to
minimize power loss in the cable.
C17*
AC power cable
C18
Synchronization
cable
If you use more than one Wide Band Transceiver (WBT) in your EK80 system, all
synchronization input signals to the Auxiliary ports must be provided by the same
source.
C19
Not used
C20**
Transducer cable
The transducer cable is provided with the transducer.
C21*
AC power cable
C22-24
Not used
C25*
AC power cable
The power cable on the Ethernet switch may comprise a separate power supply.
C26–29
Not used
EC150-3C cables
Cable
Type
C8
Ethernet cable
It is very important that a high-quality Ethernet cable is used. You must use CAT-5E
STP (Shielded Twisted Pair) quality or better. If you use a cable with lower bandwidth
capacity you will reduce the EK80 performance.
C30*
AC power cable
C31
Synchronization
cable
Synchronization is an optional feature. We strongly recommend that each signal cable
is equipped with galvanic isolation. Galvanic isolation reduces the risk of noise.
C32
Ground cable
394149/D
From/To
From Power Supply Unit to
Wide Band Transceiver (WBT)
From Power Supply Unit to AC
power outlet
From Wide Band Transceiver
(WBT) Auxiliary socket to
external synchronisation system.
From Wide Band Transceiver
(WBT) to transducer
From WBT Cabinet to AC power
outlet
From Ethernet switch to AC
power outlet
From/To
From Power Supply Unit to
Processor Unit
From Power Supply Unit to
AC power outlet
From Power Supply Unit to
synchronization system
From Power Supply Unit to
vessel ground
Cable layout and interconnections
Minimum requirements
2
2 x 1.5 mm
+ 1.5 mm
4 x 0.5 mm²
2 x 1.5 mm
2
+ 1.5 mm
2 x 1.5 mm
2
+ 1.5 mm
Minimum requirements
CAT5-E STP (Shielded Twisted
Pair)
2 x 1.5 mm
2
+ 1.5 mm
2 x 2 x 0.5 mm
2
1 x 6 mm
2
2
Ground
2
Ground
2
Ground
2
Ground
95