Changing The Prosthesis Setting Using The Cockpit App; Overview Of Adjustment Parameters In Basic Mode - Otto Bock Genium X3 3B5-3 Instructions For Use Manual

Hide thumbs Also See for Genium X3 3B5-3:
Table of Contents

Advertisement

Use
Information for changing the prosthesis settings
Before changing settings, always check the main menu of the Cockpit app to make sure the correct component
has been selected. Otherwise parameters could be changed for the wrong component.
It is not possible to change prosthesis settings nor to switch to a different mode while the prosthesis battery is
being charged. Only the status of the prosthesis can be called up. Instead of the
appears in the bottom row of the screen in the cockpit app.
The O&P professional's setting is located in the middle of the scale. After making adjustments, this setting can
be restored by tapping the "Standard" button in the Cockpit app.
Prosthesis settings should be optimised using the adjustment software. The Cockpit app is not intended for
use by the O&P professional to set up the prosthesis. The patient can use the app to change the behaviour of
the prosthesis to a certain extent during everyday use (e.g. while becoming accustomed to the prosthesis). The
O&P professional can use the adjustment software to trace these changes at the patient's next appointment.
If the settings of a MyMode are to be modified, one must first switch to this MyMode.

9.2.1 Changing the prosthesis setting using the cockpit app

Product
MyModes
Functions
Settings
Status
General
Manage components
Imprint/Info

9.2.2 Overview of adjustment parameters in basic mode

INFORMATION
When mute mode (silent mode) is activated, no beep and vibration signals are generated.
The parameters in basic mode describe the dynamic behaviour of the prosthesis in a normal gait cycle. These para­
meters act as basic settings for automatically adjusting the damping behaviour to the current motion situation (e.g.
ramps, slow walking speed, etc.).
The stance function, sitting function and/or stairs and obstacles function can also be activated/deactivated. Further
information on the stance function (see Page 34), sitting function (see Page 35), stairs and obstacles function (see
Page 36).
The following parameters can be modified:
Parameter
Resistance
Angle
38
1) Once the component is connected and in the desired mode, tap the
 symbol in the main menu.
→ The navigation menu opens.
2) Tap the menu entry "Settings".
change
→ A list appears with the parameters for the currently selected mode.
3) Change the setting of the desired parameter by tapping the "<", ">"
2. Golf
symbols.
INFORMATION: The prosthetist's setting is marked and, after the
setting has been changed, can be restored by tapping the "Stand­
ard" button.
Adjustment soft­
ware range
120 – 180
55° – 70°
Setting range,
Meaning
app
+/- 10
Resistance against flexion motion e.g. when
walking down stairs or when sitting down
+/- 3°
Maximum flexion angle in the swing phase
 symbol, the
 symbol
Genium X3 3B5-3/3B5-3=ST