7.2

Handle Alarms

7.2.1

Identify an Alarm Condition

If an alarm condition is detected, the Vivo 2 main unit and the remote alarm unit (if
connected) will alarm without delay. The alarms will remain active until the alarm condi-
tion is resolved.
Active alarms are indicated by:
Audible signal, see page 86.
Alarm message on the screen, see page 87.
The alarm LED, see page 87.
Alarm Audio signal
High priority alarms
3 short signals followed by 2 more
after 0.5 s. The signal sequence
repeats every 3rd second.
Medium priority alarms
3 signals. The signal sequence
repeats every 6th second.
For information about adjusting the alarm sound level, see 7.2.5 Adjust the Alarm Sound
Level, page 89.
Information Message Audio Signal
Short signal every 5 seconds
Alarm Sound Pressure Level
The alarm sound pressure level is adjustable within 55–80 dBA.
86
Alarms
User manual
Doc. 007229 D-3