Low Pulse Rate Alarm - Breas Vivo 2 User Manual

Hide thumbs Also See for Vivo 2:
Table of Contents

Advertisement

7.3.15

Low Pulse Rate Alarm

Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
Setting resolution
This alarm requires a connected SpO
104
Alarms
User manual
Description
Low Pulse Rate
High
A low pulse rate alarm will be given when the measured
pulse rate goes below the alarm limit for 15 seconds.
Bad positioning of the finger probe.
Too low flow of bleed-in oxygen.
Insufficient ventilatory support.
The pulse rate goes back above the alarm limit.
The Vivo 2 will continue treatment according to the current
settings.
30 to 230 bpm (beats per minute)
Off
5 bpm (beats per minute)
sensor.
2
Doc. 007229 D-3

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivo 3

Table of Contents