1 Press and hold the On/Off button
until the Starting Treatment progress bar
is filled.
⇒The Vivo 2 now performs a function
test indicated by a short beep and then
starts the treatment. Check that the self
test is performed successfully, this is indicated by:
• A short beep indicating successful alarm signalling
• The alarm LED first lights yellow, then red
• The audio pause LED lights yellow
• In about a second, both LEDs are turned off
If the function test beep absents, take the Vivo 2 out of use and contact your supplier of
the Vivo 2 .
Confirm Start/Stop is set to Off
1 Make sure that the Vivo 2 is in Standby mode.
2 Press the On/Off button.
3 The machine performs a function test as described above.
6.3

Stop the Treatment

It is possible to choose how the machine will start and stop. If you want to confirm the
starting and stopping or not, see3.5.5.1 Device Settings, page 41
Confirm Start/Stop is set to On
1 Press and hold the On/Off button
until the Stopping Treatment progress bar
is filled.
⇒A request for a complementary action
is now displayed for stopping the
treatment. If no complementary action
is taken within 6 seconds, the Vivo 2
reverts to normal operation.