6.6.7
Using the Lightweight Mobility Bag
The lightweight Mobility Bag is intended for mobile use of the Vivo 2 in hospitals, insti-
tutions and home care environments. It can be used while the ventilator is operating, for
example mounted on a wheelchair, in a personal vehicle, or carried by hand.
The lightweight Mobility Bag protects the ventilator from, water spill, sunlight, dust and
dirt, under normal handling.
The lightweight mobility bag cannot be used together with the attachable humidifier.
When using the lightweight mobility bag, the air path bypass unit shall be installed in
place of the attachable humidifier.
It does not protect against environmental impact such as shock, from rain or
snowfall.
6.6.8
Using the Y-Cable
The Y-cable is used for connecting the Vivo 2 to both mains and external DC at the
same time. see 5.3 Connecting the Vivo 2 to Power Supply, page 60. When both power sour-
ces are available, the mains will be used.
6.6.9
Using the Vivo 2 with the Trolley
Intended Use
The intended use of the trolley system is to allow patient mobility while receiving venti-
lator treatment. The trolley shall only be used in indoor, hospital environment. The trol-
ley system consists of a trolley base and a mounting bracket.
This section describes how to use the Vivo 2 and a trolley with mounting bracket.
Mount and dismount the Vivo 2 as shown in the picture:
The bottom plate is mounted to the trolley using four screws.
Before mounting the bottom plate, the attachment must be dismounted: