7.3.7
Rebreathing Alarm
Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
98
Alarms
User manual
Description
Rebreathing
High
A Rebreathing alarm will be given if the intentional leakage
is too low for more than 15 seconds.
•
Obstructed or occluded patient circuit.
•
Incorrect patient circuit.
•
Obstructed or removed CO
The leakage is back within limits.
The Vivo 2 will continue treatment according to the current
settings.
•
On
•
Off
port from leakage circuit.
2
Doc. 007229 D-3