Rebreathing Alarm - Breas Vivo 2 User Manual

Hide thumbs Also See for Vivo 2:
Table of Contents

Advertisement

7.3.7

Rebreathing Alarm

Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
98
Alarms
User manual
Description
Rebreathing
High
A Rebreathing alarm will be given if the intentional leakage
is too low for more than 15 seconds.
Obstructed or occluded patient circuit.
Incorrect patient circuit.
Obstructed or removed CO
The leakage is back within limits.
The Vivo 2 will continue treatment according to the current
settings.
On
Off
port from leakage circuit.
2
Doc. 007229 D-3