7.3.13
Low SpO
Property
Alarm text
Priority
Definition
Possible cause
Setting range
Setting resolution
Ventilator action
This alarm requires a connected SpO
102
Alarms
User manual
Alarm
2
Description
Low SpO2
High
A Low SpO
below the alarm limit for 30 seconds.
•
Too low flow of bleed-in oxygen.
•
Oxygen inlet is disconnected.
•
Delivered tidal volumes are too small.
70% to 100%
1%
The ventilator will continue treatment with the same settings.
alarm will be given when the measured SpO
2
sensor.
2
is
2
Doc. 007229 D-3