Ramp Pressure; Humidifier; Humidifier Level; Heated Circuit - Breas Vivo 2 User Manual

Hide thumbs Also See for Vivo 2:
Table of Contents

Advertisement

4.2.4

Ramp Pressure

The ramp pressure setting is used for defining the start and end pressure when Ramp
Up and/or Ramp Down is configured.
IPAP shall start at 2 cmH
is set below EPAP, the EPAP pressure will start at the configured Ramp Pressure.
CPAP mode:
Unit
cmH
O
2
4.2.5

Humidifier

This setting defines whether the attachable humidifier shall be used.
WARNING!
Read the section 2.6 Humidification and Heating — Warnings and Precautions, page 24
before activating the humidifier, to make sure all conditions are met and considered.
Min
Off
4.2.6

Humidifier Level

This setting defines the level of humidification, if Humidifier is set to On.
Min
1
4.2.7

Heated Circuit

This setting defines whether a heated patient circuit shall be used as patient circuit.
WARNING!
Read the section 2.6 Humidification and Heating — Warnings and Precautions, page 24
before activating the heated circuit, to make sure all conditions are met and
considered.
Doc. 007229 D-3
O above the configured Ramp Pressure. If the Ramp Pressure
2
Min
Max
2
EPAP or
CPAP
Max
On
Max
5
Default
Resolution
2
0.5 < 10
1.0 ≥ 10
Treatment Functions and Settings
Tolerance
± (0.5 cmH
0 + 5%)
2
Default
Off
Default
1
User manual
57