2 Do one of the following:
• Press the Mute Alarm button
(a). (Always available.)
• Press Ramp (b) for a ramped
stop. (Available if a Ramp Down
time has been specified.)
Confirm Start/Stop is set to Off
1 Press the On/Off button.
2 Press Ramp (b) for a ramped stop. (Available if a Ramp Down time has been
specified.)
6.4
Switch Off the Vivo 2
Confirm Start/Stop is set to On
1 Make sure that the treatment is stopped and the Vivo 2 is in Standby mode.
2 Press the On/Off button.
⇒When the message "Do you want to
turn off the ventilator?" is displayed,
confirm by pressing the Mute Alarm
button within 6 seconds. Otherwise, the
Vivo 2 will revert to standby mode.
66
How to Use the Vivo 2
User manual
a
Doc. 007229 D-3