Using The Remote Alarm Unit - Breas Vivo 2 User Manual

Hide thumbs Also See for Vivo 2:
Table of Contents

Advertisement

Supplemental Oxygen Supply Requirements
Property
Maximum flow
Maximum pressure
Supply source
Connector
Connect the Oxygen supply
1 Connect the oxygen adapter (article no. 005032) to the oxygen supply's tube.
2 Connect the oxygen adapter to the oxygen port at the back of the Vivo 2. See 3.3
Ventilator Back , page 33 for detailed information.
3 If using a device for monitoring the oxygen saturation, connect it according to the
supplier instructions.
CAUTION!
Do not add Oxygen before therapy has been started.
6.6.5

Using the Remote Alarm Unit

Information about safety, warnings, product description, installation, usage, cleaning,
maintenance and technical specifications can be found in the user instruction for
Remote Alarm.
The Remote Alarm enables care providers and clinical personnel to monitor the Vivo 2
alarms remotely. The Remote Alarm forwards alarms from the Vivo 2.
1 Connect the Accessory box to the Communication port at the back of the Vivo 2.
80
How to Use the Vivo 2
User manual
Requirement
30 l/min.
100 kPa
Source equipped with rotameter.
Examples of supply sources:
Oxygen cylinder
Central oxygen supply
Oxygen concentrator
The oxygen source shall be equipped
with the Breas Low Pressure Oxygen
Adapter, art. no. 005032
Doc. 007229 D-3

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivo 3

Table of Contents