Using The Oxygen Supply Adapter - Breas Vivo 2 User Manual

Hide thumbs Also See for Vivo 2:
Table of Contents

Advertisement

CAUTION!
6.6.4

Using the Oxygen Supply Adapter

WARNING!
Read the section 2.8 Oxygen Usage — Warning and Precautions, page 27 before using the
Vivo 2 with oxygen.
Supplemental oxygen with a flow up to 30 l/min can be added by an oxygen source
with rotameter such as oxygen cylinder, central oxygen supply system or an oxygen
concentrator.
Doc. 007229 D-3
When using the Vivo 2 with the SpO2 sensor, the Vivo 2
displays functional oxygen saturation measured by the sensor.
The following information concerns the light emitted by the SpO2
Peak Wavelength (red): 660 nm
Peak Wavelength (infrared): 905 nm
Maximum Optical Output Power: ≤ 15 mW
For more information regarding the oxygen probe's range of peak
wavelengths, max optical power and usage, please refer to the respec-
tive probe manual.
Environmental factors may influence the function or accuracy of the pulse oxi-
meter, such as ambient light, physical movement, diagnostic testing, low perfusion,
electromagnetic interference, dysfunctional haemoglobin, presence of certain dyes
and inappropriate positioning of the pulse oximeter probe.
A functional test cannot be used to assess the accuracy of a pulse oximeter probe
or a pulse oximeter.
How to Use the Vivo 2
User manual
79