Environmental Condition
Humidity
Ambient Pressure Range
Ventilator ingress protection
9.4

Pneumatic Diagram

1
Doc. 007229 D-3
Specification
RH from 15% to 95%, non-condensing.
70 to 106 kPa
This corresponds to ~ 315 m below sea level to
~3000 m above sea level
IP 22
Mechanical ingress protection: protected from
touch by fingers and objects greater than 12.5
mm
Liquid ingress protection: The device withstands
dripping water(equivalent to 3 mm rainfall
/minute) when not tilted more than 15 degrees
from vertical. The protection has been tested for
10 minutes (2.5 minutes in every tilt direction).
Precautions
There is a silicone lid to protect the USB, SD card,
and communication ports. The IP22 classification
is applicable only when this lid is in place. How-
ever, the accessory box can be connected with
retained IP22 classification, but then only the
lower part of the silicone lid can be opened.
Ensure that the silicone lid on the back of the ven-
tilator is closed when no accessories are connected.
2
3