Environmental Condition
Humidity
Ambient Pressure Range
Ventilator ingress protection
9.4
Pneumatic Diagram
1
Doc. 007229 D-3
Specification
RH from 15% to 95%, non-condensing.
70 to 106 kPa
This corresponds to ~ 315 m below sea level to
~3000 m above sea level
IP 22
Mechanical ingress protection: protected from
touch by fingers and objects greater than 12.5
mm
Liquid ingress protection: The device withstands
dripping water(equivalent to 3 mm rainfall
/minute) when not tilted more than 15 degrees
from vertical. The protection has been tested for
10 minutes (2.5 minutes in every tilt direction).
Precautions
•
There is a silicone lid to protect the USB, SD card,
and communication ports. The IP22 classification
is applicable only when this lid is in place. How-
ever, the accessory box can be connected with
retained IP22 classification, but then only the
lower part of the silicone lid can be opened.
•
Ensure that the silicone lid on the back of the ven-
tilator is closed when no accessories are connected.
2
3
Technical Specifications
User manual
4
115