Air Pathway Disinfection; Clean The Patient Circuit; Clean And Replace Patient Air Inlet Filters - Breas Vivo 2 User Manual

Hide thumbs Also See for Vivo 2:
Table of Contents

Advertisement

5 When the equipment is clean and dry, reconnect the patient circuit and any
accessories that was disconnected during the cleaning.
8.1.2

Air Pathway Disinfection

The table below lists the parts that might get contaminated by exhaled gases or bodily
fluids during normal use or single fault condition.
Condition
With bacteria filter
Without bacteria filter
In case of contamination, the internal air pathways of the Vivo 2 may be disinfected up
to 5 times by a maximum 60 minute long validated ozone gas process.
Low resistance bacteria filter, if used, should be replaced every 24 hours.
8.1.3

Clean the Patient Circuit

CAUTION!
The cleaning and replacement intervals should be established by the care provider,
based on the care provider's infection control procedures and the instructions from
the patient circuit's manufacturer.
Check the patient circuit regularly for damage. In case of damage, replace the circuit.
The patient circuit should be cleaned and replaced in accordance with the manufac-
turer's instructions, or by the care provider's instructions if additional cleaning or
other replacement intervals are prescribed.
8.2

Clean and Replace Patient Air Inlet Filters

The Vivo 2 patient air inlet filters are located inside a magnetic filter holder at the back
of the ventilator. The table below describes the filters and their minimum maintenance
intervals.
110
Cleaning and Maintenance
User manual
Parts
Patient circuit
Bacteria filter
Patient circuit
Patient air outlet
Air bypass unit/humidifier
Blower unit
Air inlet filters and filter holder
Doc. 007229 D-3