Low Epap Alarm; High Spo 2 Alarm - Breas Vivo 2 User Manual

Hide thumbs Also See for Vivo 2:
Table of Contents

Advertisement

7.3.11

Low EPAP Alarm

Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
7.3.12
High SpO
Property
Alarm text
Priority
Alarm condition
Possible cause
Reset criteria
Ventilator action
Setting range
Setting resolution
This alarm requires a connected SpO
Doc. 007229 D-3
Description
Low EPAP
Medium
A Low EPAP alarm will be given when EPAP has gone 30%
below the set value for 3 breaths.
Excessive leakage.
EPAP has gone above the alarm limit (higher than 30%
below the set value).
The Vivo 2 will continue treatment according to the current
settings.
On
Off
Alarm
2
Description
High SpO2
Medium
A High SpO
exceeds the alarm limit for 30 seconds.
Too high flow of bleed-in oxygen.
The SpO
The Vivo 2 will continue treatment according to the current
settings.
80 % to 100 %
Off
1 %
alarm will be given when the measured SpO
2
value goes back below the alarm limit.
2
sensor.
2
2
Alarms
101
User manual