How to perform a single sweep while in Hold
 Press Sweep 3.
 Then choose one of the following:

Preset

The Preset key restores the FieldFox measurement settings to a known condition.
You can preset the entire instrument, or just the current Mode (CAT, Power Meter,
NA.)
When Preset, the FieldFox becomes CalReady at the test ports. Learn more on page
63.
The following settings always survive Preset and/or Mode Preset:
 External Reference state (page 139);
 Marker Peak Excursion and Threshold (page 117).
 VVM Display Resolution (page 104).
 Volume Control
How to Preset the FieldFox
 Press Preset .
 Then choose one of the following:

User Preset

User Preset allows you to store custom settings for each mode, then recall the
settings when the Preset key is pressed. Your custom settings are saved to a
standard State file (UserPreset.sta). However, unlike State files, calibration data is
NOT saved. Learn more about State files on page 126.
How to set User Preset
 Make your custom settings for all modes.
 When finished, press Preset .
 Then
 Then
How to recall User Preset
 Press Preset .
132
-->
is displayed while a single sweep occurs.
Automatically sets Continuous OFF (Hold) and causes FieldFox to make
Single
o
ONE measurement sweep, then hold for the next Single key press.
Makes continuous sweeps. This is the typical setting when
Continuous
o
battery power is not critical.
Returns ALL Modes to the factory default settings.
Preset
o
Returns only the current Mode to the factory default settings.
Mode Preset
o
The current settings for all other Modes remain.
Preset, then select
Save User
User Preset ON
to confirm.