This page summarizes the Mechanical Cal to be performed and allows you to make
changes. For best results, review the screen and make changes in the following
order:
1. S-parameters – Verify that these are all of the S-parameters that you want
2. DUT Connectors – For each DUT port that is shown, verify the correct type and
3. Cal Kit - Verify the correct Cal Kit for each DUT port to be calibrated. See a list of
4. Calibration Type - The FieldFox always RECOMMENDS a Cal Type that will
Calibration for NA, CAT, and VVM Modes
calibrated. If not, press Esc to terminate the calibration process, then return to
the Trace menu to display all of the traces to be calibrated. Learn how on page
23.
gender of the DUT connector. If a DUT port is not shown, then that port is not
included in the list of S-parameters to be calibrated.
supported Cal Kits in the FieldFox data sheet. See page 162.
If the DUT connector type, gender, or Cal Kit is NOT correct for the DUT ports to
be calibrated, then:
Press
Change DUT Connectors
o
on your DUT.
For each port:
o
Use the ▲|▼ arrows or rotary knob to change the DUT connector type.
o
Press
Change Gender
o
Then press
Next...
o
For each port:
o
Use the ▲|▼ arrows or rotary knob to select the Cal Kit that you have, and
o
are going to use, for the specified ports.
Then press
Next...
o
quickly and accurately calibrate all of the displayed S-parameters. Change from
the RECOMMENDED Cal Type ONLY if you understand the implications. Learn
more about Cal Types on page 70.
To select a different Cal Type:
Press
Change Cal
Type.
o
Then using the ▲|▼ arrows or rotary knob, select a Cal Type,
o
to select the correct connector types that are
to change the gender of the DUT connector.
to make the selections.
to make the selection.