Keysight N9923A FieldFox User Manual page 34

Hide thumbs Also See for N9923A FieldFox:
Table of Contents

Advertisement

How to create Big Marker Display States
To have big readouts, markers MUST be created using the following procedure.
If you have already set up your display and do not want to lose it, then save the
current state to a state file. Learn how on page 124. The following procedure will
overwrite your display state.
 Press Mkr ->/ Tools
 With A or B selected, then press
 Then edit the following display state settings:
 Num Traces – Choose the Multi-Trace configuration x1, x2, or x3. Only Overlayed
 Font Size – Choose either Big or Super Big. Super Big marker readout #3 will cover
 Trace Settings - For each possible trace (1, 2, and 3) select from the following:
 Big Readout - For each marker readout (1, 2, and 3), select from the following:
Notes:
 To move a marker, press Marker, then Marker, then scroll to the marker number to
34
Then
Big Readout
o
A or B - The A or B display state is recalled. If none have been defined, then the
o
default display state is recalled.
OFF - The B display state is visible but without the Big Marker Readout.
o
configurations are allowed. Learn more on page 23.
some of the measurement grid.
Measurement – S-Parameter or Receiver measurement to display. S11 is
o
default. Learn more starting on page 21.
Meas Format – Log is default. Learn more on page 26
o
Bandwidth (ON) – Marker readout for marker bandwidth search. OFF is default.
o
Learn how to set BW parameters on page Error! Reference source not
found.114.
Marker # - Enter a number from 1 to 6 to create the marker. 1 is created by
o
default.
Trace # - The marker# is created on this trace. Tr1 is selected by default. The
o
corresponding Num Traces must first be selected. For example, to create a
marker on Trace 3, first select Num Traces = x3.
Format – Format for the marker readout. Select Default to make the marker
o
readout format the same as the display format. Default is selected by default.
Mkr Tracking – ON: Select the parameter to track on the standard Marker
o
Search menu. OFF is selected by default. When set to ON, Peak tracking is set
by default. Learn about Marker Tracking on page 114.
State – Choose from Normal, Delta, or OFF. The default marker (1) is created in
o
Normal. Learn more on page 111.
Enabled – Turns the Big Readout of the corresponding marker ON and OFF. The
o
default marker (1) is set ON by default.
be moved. The new marker position can NOT be saved with the A or B display
state.
Edit Big Marker (A or B)
FieldFox User's Guide

Advertisement

Table of Contents
loading

Table of Contents