Special Options
Option SF2
— Save
N O T E
Save and Recall are often used with the N9069C Noise Figure Measurement
Application to allow multiple noise sources to be used without the need to
re-enter ENR values or to perform the noise source calibration each time a
different noise source is connected. When Option SF2 is installed, this
functionality is not available.
File Menu
The Page Setup, Print, Quick Save, Recall and Save icons are not available.
Preset Menu
The User Preset, Save User Preset, and User Preset All Modes icons are not
available.
System Menu
In the System > Power On menu, the Power On Type selection keys (Mode
Preset, User Preset and Last State) are not available.
Disabled SCPI Commands
The following SCPI commands are not available. Attempting to use any of
these commands generates the Error message:
Note that some of the commands listed are specific to individual applications,
thus the set of available commands depends on which applications are
licensed for a given instrument.
IEEE 488.2 Common Commands (*)
—
—
MMEMory Subsystem
—
—
—
—
—
—
—
—
—
Security Features & Document of Volatility
*RCL
*SAV
:MMEMory:CLEar:ALL
:MMEMory:CLEar:STATe
:MMEMory:COMMent
:MMEMory:COPY:DEVice
:MMEMory:INITialize
:MMEMory:LOAD:ALIMit
:MMEMory:LOAD:ALISt
:MMEMory:LOAD:ATRace
:MMEMory:LOAD:AUTO
–113,"Undefined
header".
70