Normalize uses a THRU standard or cable between port 1 and port 2 to cal an S21
and S12 Transmission measurement (NA Mode) and a 2-port Insertion Loss
measurement (CAT Mode). In VVM Mode, this is performed using Zero.
How to perform a Response Cal
 Select the measurements to be calibrated. See the relevant Mode (NA, CAT, or
 Press Cal 5 then
 For 1-port measurements:
 For 2-port measurements:
NOTE
With a Normalization Cal, all subsequent insertion loss measurements are made
relative to the insertion loss of the THRU cable. For example, if you use a cable with 1
dB of loss, then after Normalization, the display will show 0 dB of loss with this cable
in place. Therefore, for highest accuracy, when measuring the DUT also attach the
cable that was used in the normalization cal.
 Press Finish.
 Connect the DUT.
CAL ON U
appropriate measurement.

View Cal

From the Choose Calibration screen (page 63), press
screen:
The top box shows the properties of the current calibration that you performed and
the displayed S-parameters that it is correcting.
70
VVM) for measurement selections.
Response Cal
Select either
Open Response
o
calibrated. Port 2 is available only with Option 122.
Connect an OPEN or SHORT standard to the specified port and press
o
Select
Normalization
o
Connect a short, high-quality, phase stable cable between the FieldFox port 1
o
and port 2 connectors, then press
is shown on the screen when a Response Cal is correcting ONLY the
or
for the port to be
Short Response
Measure
View Cal
Measure
to see the following