5.8 Checking the calibration
Live view
Ø 10 mm
Ø 10 mm
0.0 Nm
0.0 %
> 5 mm
> 5 mm
5.9 Switching off
120
0.0 Nm
0.0 %
1
2
±5 mm
±5 mm
> 5 mm
> 5 mm
3D L.A.S.A.R.
Posture
Turn off
1
2
Calibration in this section serves to check the measur
ing accuracy of the product against specified toler
ances.
Test interval: 1x per year
>
Test equipment: gymnastics rod with rounded
>
ends (Ø 10 mm)
>
Mode: 3D mode activated (green)
► With your hand flat, move the gymnastics rod to dif
ferent positions.
Pause at each position without sliding the gym
nastics rod on the force measurement plate, thereby
generating additional moments.
Check the load line on the gymnastics rod while
pausing.
→ The load line has to follow the longitudinal axis
of the gymnastics rod and may not deviate more
than ±5 mm from the marked height of 500 mm,
which corresponds to the edge of the gym
nastics rod.
→ If the deviation is > 5 mm, the force measurement
plate has to be sent to the manufacturer for service.
► NOTICE! Send the 3D L.A.S.A.R. Posture to the
manufacturer every two years to have the calib
ration checked.
Tablet
► Press the on/off button of the tablet for 2 seconds.
► Confirm the shutdown message.
→ The tablet is turned off.
Mini computer
► Pull the plug of the mini computer's power supply
out of the outlet.
→ The mini computer is switched off.