Use

5.5.7 Camera connection failed

5.5.8 Use outside of the permissible temperature range

106
Error message text:
: "Camera connection failed"
: "Please check the cables between the PC and the
cameras"
: "OK"
Possible causes:
USB plug not connected to mini-computer
Mini-computer software has crashed
Defective cable
Defective camera
Measures:
► Carry out the appropriate measures depending on
the cause (for example connect USB plug; restart
the mini-computer; have defective components
repaired by Ottobock Service).
► 1x tap on "OK".
Error message text:
: "Use outside of the permissible temperature range"
: "The operating temperature is outside the permiss­
ible temperature range."
: "Measurement errors may occur."
: "OK"
Possible causes:
The ambient temperature is < +15 °C.
The ambient temperature is > +30 °C.
Measures:
► Use the product exclusively within the permissible
temperature range (+15 °C – +30 °C).
► Tap on "OK" once.