Technical data
Component at risk of electrostatic discharge
Non-ionising radiation
Medical device
10 Technical data
Supply voltage (mini PC)
Supply voltage (tablet battery charger)
Weight (3D L.A.S.A.R. Posture, complete in case)
Case dimensions
Operating temperature
Dust and water resistance (applies only to the force measurement
plate)
Max. body weight
Measuring tolerances
Centre of pressure (COP) at the bottom end of the load line
Relative body weight
Directives and manufacturer's declaration – electromagnetic interference
The 743L500 3D L.A.S.A.R. Posture is designed for operation in an electromagnetic environment as specified
below. The customer or user of the 743L500 3D L.A.S.A.R. Posture must ensure that the device is operated in
such an environment.
Interference measurements
HF emissions according to CIS
PR11
HF emissions according to CIS
PR11
Emission of harmonics according to
IEC 61000-3-2
Emission of voltage fluctuations/flick
er according to IEC 61000-3-3
Directives and manufacturer's declaration – electromagnetic interference immunity
The 743L500 3D L.A.S.A.R. Posture is intended for operation in the electromagnetic environment specified
below. The customer or user of the 743L500 3D L.A.S.A.R. Posture must ensure that the device is operated in
such an environment.
INTERFERENCE
IMMUNITY TESTS
126
Compliance
Group 1
Class B
Class D
Complies
IEC 60601 test level
110–240 V AC/50–60 Hz
110–240 V AC/50–60 Hz
20 kg
280 x 630 x 520 mm
15°–30° C
IP21 (protection against penetration of solid
foreign objects with a diameter > 12.5 mm,
protection against vertically dripping water)
150 kg
The following tolerances refer to the camera
view shown on the tablet.
± 1 mm (left-right alignment)
± 5 mm (depth alignment)
± 3% of the measured value
Electromagnetic
directives
The 743L500 3D L.A.S.A.R. Posture
uses HF energy exclusively for its
internal FUNCTIONING. Therefore
its HF emissions are very low and
interference with neighbouring elec
tronic devices is unlikely.
The 743L500 3D L.A.S.A.R. Posture
is intended for use in all facilities
including residential, and those con
nected directly to the PUBLIC SUP
PLY NETWORK which also supplies
buildings used as residences.
Compliance level
environment
ELECTROMAGNETIC
ENVIRONMENT – Direct
ives
743L500 3D L.A.S.A.R. Posture
-