5.5.5 Shift detected
5.5.6 Warning
743L500 3D L.A.S.A.R. Posture
Error message text:
: "Shift detected"
: "Platform shift detected"
: "Cancel"
: "Recalibrate"
Possible causes:
•
The camera stand or force measurement plate was
moved
•
All LEDs were pushed out of position
•
Underexposed image (black)
Measures:
► Carry out the appropriate measures depending on
the cause (for example align the force measurement
plate and camera stands so that all LEDs are recog
nised within the outer blue frame; restart the mini-
computer in case of incorrect exposure).
► 1x tap on "Recalibrate".
Error message text:
: "Warning"
: "Measuring accuracy too low."
: "Realign the measurement plate." – Position the
force measurement plate closer to the camera stand to
use the full camera resolution.
: "Cancel"
: "Recalibrate"
Possible causes:
•
Force measurement plate rotated too much relative
to the camera
•
Camera stand tilted away from the force measure
ment plate too much
•
Camera too far away from the force measurement
plate
Measures:
► Carry out the appropriate measures depending on
the cause (for example align the force measurement
plate and camera stands so that all LEDs are recog
nised within the outer blue frame, that the
crosshairs of the inner blue frame are centred over
the middle of the force measurement plate, and that
the hinges on the force measurement plate are
aligned with one of the two centre lines of the
crosshairs; align the camera at a 90° angle; reduce
the distance between the camera and force meas
urement plate).
► 1x tap on "Recalibrate".
Use
105