5.5.9 Defective platform
5.5.10 Unfortunately, the 3D L.A.S.A.R. Posture app has stopped.
743L500 3D L.A.S.A.R. Posture
Error message text:
: "Defective platform"
: "The platform is defective."
: "Please contact your local customer service."
: "OK"
Possible causes:
•
Broken cable
•
Torn cable
•
Defective circuit board
Measures:
► Have the defective force measurement plate
repaired by Ottobock Service.
► 1x tap on "OK".
Error message text:
: "Unfortunately, the 3D L.A.S.A.R. Posture app has
stopped."
: "OK"
Possible causes:
•
Unexpected app error
Measures:
► 1x tap on "OK" and restart the app.
► Contact Ottobock Service if this problem recurs
often.
Use