Use

5.8 Checking the calibration

Live view
Ø 10 mm
Ø 10 mm
0.0 Nm
0.0 %
> 5 mm
> 5 mm

5.9 Switching off

56
0.0 Nm
0.0 %
1
2
±5 mm
±5 mm
> 5 mm
> 5 mm
3D L.A.S.A.R.
Posture
Turn off
1
2
Calibration in this section serves to check the measur­
ing accuracy of the product against specified toler­
ances.
Test interval: 1x per year
>
Test equipment: gymnastics rod with rounded
>
ends (Ø 10 mm)
>
Mode: 3D mode activated (green)
► With your hand flat, move the gymnastics rod to dif­
ferent positions.
Pause at each position without sliding the gym­
nastics rod on the force measurement plate, thereby
generating additional moments.
Check the load line on the gymnastics rod while
pausing.
→ The load line has to follow the longitudinal axis
of the gymnastics rod and may not deviate more
than ±5 mm from the marked height of 500 mm,
which corresponds to the edge of the gym­
nastics rod.
→ If the deviation is > 5 mm, the force measurement
plate has to be sent to the manufacturer for service.
► NOTICE! Send the 3D L.A.S.A.R. Posture to the
manufacturer every two years to have the calib­
ration checked.
Tablet
► Press the on/off button of the tablet for 2 seconds.
► Confirm the shutdown message.
→ The tablet is turned off.
Mini computer
► Pull the plug of the mini computer's power supply
out of the outlet.
→ The mini computer is switched off.