AMI Turbitrace
Installation
3.6.
Installation
requirements
22
Power Supply
WARNING
Risk of electrical shock
Installation and maintenance of electrical parts must be per-
formed by professionals
Always turn off AC power before manipulating electric parts.
A
B
C
D
A
Power supply connector
B
Neutral conductor, Terminal 2
C
Phase conductor, Terminal 1
D
Protective earth PE
NOTICE: The protective earth wire (Ground) has to be
connected to the grounding terminal.
The installation must meet the following requirements.
Mains fuse 1.6 AT
Mains cable to comply with standards IEC 60227 or IEC
60245; flammable rating FV1
Mains equipped with an external switch or circuit-breaker
– near the instrument
– easily accessible to the operator
– marked as interrupter for AMI Turbitrace
A-96.250.131 / 050220