Maintenance Schedule

Preventive maintenance frequency depends on water quality, on
the application, and on national regulations.
Every week
If necessary
When maintenance
message E068
appears
Check sample flow.
Clean the measuring chamber.
Replace the 0-point filter
Replace the humidity absorber.
Recommendation:
When replacing the humidity absorber,
check also the O-ring of the bayonet lock
cover and replace it if necessary.